Bently Nevada TK3 Proximity System Test Kit
Buy Bently Nevada TK3 in Dubai, UAE at unbeatable prices! Explore our range of high-quality products for reliable solutions.
- In-stock items are delivered within 2 business days.
- We are not Supporting COD Payments Terms Outside UAE
- For out-of-stock items, delivery may take 2 to 4 business weeks.
Order Via phone number
Order Via Online Form
Bently Nevada TK3 Proximity System Test Kit
The TK-3 Proximity System Test Kit is designed to simulate shaft vibration and position for the calibration of Bently Nevada monitors. It checks both the operational status of the monitor readouts and the condition of the proximity transducer system. Proper calibration of the system guarantees the accuracy of the transducer inputs and the corresponding monitor readings.
The TK3 vibration calibrator employs a detachable spindle micrometer assembly to verify the calibration of the transducer system and position monitor. This assembly includes a universal probe mount that can accommodate probe diameters ranging from 5 mm to 19 mm (0.197 in to 0.75 in). The mount secures the probe while the user adjusts the target distance to or from the probe tip in precise increments, recording the output from the Proximity Sensor with a voltmeter. Additionally, the spindle micrometer assembly is equipped with a magnetic base for easy field use.
Bently Nevada TK3 Key Features
- Light weight and portable
- Maintenance-free electric drive powered
- Industrial-strength case dramatically improves durability and better protects the precision devices inside
- Detachable spindle micrometer with magnetic base
- Universal probe mount accommodates probe diameters from 5 to 19mm
Bently Nevada TK3 Specifications
| Category | Specification |
| Power Requirements | Electric: 95-125 Vac, 50/60 Hz, 1A minimum <br> 190-250 Vac, 50/60 Hz, 1A minimum <br> Power can be disconnected by removing the power cable from the TK-3 unit. |
| Air: 90 psi (6.2 bar) maximum | |
| Wobulator Range | Vibration Amplitude Range: 50 μm to 254 μm (2 to 10 mils) peak-to-peak |
| Maximum Speed | Electric: 0 to 5000 cpm ± 1000 cpm <br> Air: 0 to 5000 cpm ± 1000 cpm |
| Spindle Micrometer Range | 0 – 25.4 mm (0 – 1000 mils) |
| Target Button and Wobble Plate | AISI 4140 Alloy Steel <br> Contact your nearest Sales Professional for details on special target and wobble plate materials. |
| Physical Size | Height: 195 mm (7.68 inches) <br> Width: 299 mm (11.8 inches) <br> Depth: 248 mm (9.76 inches) <br> Weight: 5.22 kg (11.5 lb) |
| Environmental | Equipment is for indoor use only. <br> Maximum altitude is 2000 m. |
| Operational Temperature Range | 0 °C to 50 °C (32 °F to 122 °F) <br> Temporary operation below minimum temperature is acceptable. <br> For extended continuous use, maximum operating temperature is 40°C. |
| Storage Temperature Range | -18 °C to 65 °C (0 °F to 150 °F) |
| Humidity | 95% Non-Condensing Humidity |
| Exposure | Designed to meet IP54 for dust and water exposure (closed) |
| velocity__range (vibration meter) |
none |
|---|---|
| acceleration__range (vibration meter) |
none |
| axis__count (vibration meter) |
none |
| frequency___range (vibration meter) |
none |
Specifications
Bently Nevada TK3 Specifications
| Category | Specification |
| Power Requirements | Electric: 95-125 Vac, 50/60 Hz, 1A minimum <br> 190-250 Vac, 50/60 Hz, 1A minimum <br> Power can be disconnected by removing the power cable from the TK-3 unit. |
| Air: 90 psi (6.2 bar) maximum | |
| Wobulator Range | Vibration Amplitude Range: 50 μm to 254 μm (2 to 10 mils) peak-to-peak |
| Maximum Speed | Electric: 0 to 5000 cpm ± 1000 cpm <br> Air: 0 to 5000 cpm ± 1000 cpm |
| Spindle Micrometer Range | 0 – 25.4 mm (0 – 1000 mils) |
| Target Button and Wobble Plate | AISI 4140 Alloy Steel <br> Contact your nearest Sales Professional for details on special target and wobble plate materials. |
| Physical Size | Height: 195 mm (7.68 inches) <br> Width: 299 mm (11.8 inches) <br> Depth: 248 mm (9.76 inches) <br> Weight: 5.22 kg (11.5 lb) |
| Environmental | Equipment is for indoor use only. <br> Maximum altitude is 2000 m. |
| Operational Temperature Range | 0 °C to 50 °C (32 °F to 122 °F) <br> Temporary operation below minimum temperature is acceptable. <br> For extended continuous use, maximum operating temperature is 40°C. |
| Storage Temperature Range | -18 °C to 65 °C (0 °F to 150 °F) |
| Humidity | 95% Non-Condensing Humidity |
| Exposure | Designed to meet IP54 for dust and water exposure (closed) |