Eurotron PLB 157 Micro Bath Temperature Calibrator

Short Description:

Order Eurotron products from us at competitive prices, available across the UAE and KSA.

Product purchase
SKU: 87262
Brand:Eurotron

Call for Best Price 00971(04) 447 6031 00971(04) 344 9868

16 People watching this product now!
Terms of sending goods ›
  •  In-stock items are delivered within 2 business days.
  • We are not Supporting COD Payments Terms Outside UAE
Description

Eurotron PLB 157 Micro Bath Temperature Calibrator

The Eurotron PLB 157 Portable Liquid Bath is designed for precise calibration of temperature sensors, offering a high degree of accuracy and versatility in a compact and portable form. This liquid bath is ideal for on-site calibrations and laboratory use, providing superior performance and reliability.

The Eurotron PLB 157 Temperature Bath Calibrator offers a robust and versatile solution for temperature sensor calibration, combining portability with high performance. Its user-friendly design and superior accuracy make it a valuable asset for professionals requiring reliable and precise calibration in various settings.

The PLB 157 is available in two versions- the TS and TC.  The TS  has a digital display for set and nominal temperature, the TC additionally includes inputs for reference sensor and test thermometers.

Eurotron PLB 157 Dry Bath Temperature Calibrator Key Features

  • Large Calibration Volume
  • 4 operating modes
  • Use as Liquid bath
  • Use as Dry Block with optional inserts
  • Use as Surface Temperature calibrator with optional insert
  • Use a Infrared Sensor Calibrator with optional insert
  • Extended features including full programmable auto ramp
  • Dual Inputs for reference and sensor under-test display ( TC version)

Eurotron PLB 157 Specifications

Temperature Range
@ 20ºC -38C to 160 C
@ 0ºC -50 to 160°C
Stability
As Liquid Bath ± 0.01 °C
As Dry block ± 0.01 °C
Uniformity as Liquid Bath
Axial ± 0.02 °C @ -10 °C
± 0.02 °C@ 0°C
± 0.01 °C @ 120 °C
Radial ± 0.01 °C @ -10 °C
± 0.01 °C@ 0°C
± 0.02 °C @ 120 °C
Accuracy
with Internal Sensor ± 0.15 °C
with External Sensor SRS (*) ± 0.08 °C   ( TC version only)
Display 
Set Point -50 to 160°C
Resolution user-selectable 0.1 , 0.01 C
Units user-selectable C, F,K
Calibration Volume
As Liquid Bath 44.5 x 185 mm
As dry block 45 x 170 mm ( dry inserts optional)
Heating time ambient to 120°C: 25 min ( include time to stability)
Cooling time ambient to -10°C : 25 min (include time to stability)
Thermocouple Input (TC version)
Type K,J,N,R,S,E,T,B
Range 0 to 1300 °C
Accuracy ± 1°C Include CJC error
Resolution user-selectable 0.1 , 0.01 °C
RTD’S input ( TC version only)
Type Pt 100 (3/4 wires)
Range -100 C to 700 °C
Accuracy ± 0.3 °C
Resolution user-selectable 0.1 , 0.01 °C
General
Computer interface RS 232 as standard ( UBS adapter supplied as option) and EIUK software included
Switch Test Yes
Ramp Yes fully programmable
Weight 10.2 Kg
Dimensions 160x3350x365 mm
Power Requirement 115 o 230 V – 50/60 Hz
Power 300 VA
 Notes * Require SRS sensor and special UKAS matching calibration. Long term drift not included
Datasheet

Datasheet

Mme-Pdfs-Image

Specifications

Specifications

Eurotron PLB 157 Specifications

Temperature Range
@ 20ºC -38C to 160 C
@ 0ºC -50 to 160°C
Stability
As Liquid Bath ± 0.01 °C
As Dry block ± 0.01 °C
Uniformity as Liquid Bath
Axial ± 0.02 °C @ -10 °C
± 0.02 °C@ 0°C
± 0.01 °C @ 120 °C
Radial ± 0.01 °C @ -10 °C
± 0.01 °C@ 0°C
± 0.02 °C @ 120 °C
Accuracy
with Internal Sensor ± 0.15 °C
with External Sensor SRS (*) ± 0.08 °C   ( TC version only)
Display 
Set Point -50 to 160°C
Resolution user-selectable 0.1 , 0.01 C
Units user-selectable C, F,K
Calibration Volume
As Liquid Bath 44.5 x 185 mm
As dry block 45 x 170 mm ( dry inserts optional)
Heating time ambient to 120°C: 25 min ( include time to stability)
Cooling time ambient to -10°C : 25 min (include time to stability)
Thermocouple Input (TC version)
Type K,J,N,R,S,E,T,B
Range 0 to 1300 °C
Accuracy ± 1°C Include CJC error
Resolution user-selectable 0.1 , 0.01 °C
RTD’S input ( TC version only)
Type Pt 100 (3/4 wires)
Range -100 C to 700 °C
Accuracy ± 0.3 °C
Resolution user-selectable 0.1 , 0.01 °C
General
Computer interface RS 232 as standard ( UBS adapter supplied as option) and EIUK software included
Switch Test Yes
Ramp Yes fully programmable
Weight 10.2 Kg
Dimensions 160x3350x365 mm
Power Requirement 115 o 230 V – 50/60 Hz
Power 300 VA
 Notes * Require SRS sensor and special UKAS matching calibration. Long term drift not included