Eurotron PLB 157 Micro Bath Temperature Calibrator
The Eurotron PLB 157 Portable Liquid Bath is designed for precise calibration of temperature sensors, offering a high degree of accuracy and versatility in a compact and portable form. This liquid bath is ideal for on-site calibrations and laboratory use, providing superior performance and reliability.
The Eurotron PLB 157 Temperature Bath Calibrator offers a robust and versatile solution for temperature sensor calibration, combining portability with high performance. Its user-friendly design and superior accuracy make it a valuable asset for professionals requiring reliable and precise calibration in various settings.
The PLB 157 is available in two versions- the TS and TC. The TS  has a digital display for set and nominal temperature, the TC additionally includes inputs for reference sensor and test thermometers.
Eurotron PLB 157 Dry Bath Temperature Calibrator Key Features
- Large Calibration Volume
- 4 operating modes
- Use as Liquid bath
- Use as Dry Block with optional inserts
- Use as Surface Temperature calibrator with optional insert
- Use a Infrared Sensor Calibrator with optional insert
- Extended features including full programmable auto ramp
- Dual Inputs for reference and sensor under-test display ( TC version)
Eurotron PLB 157 Specifications
Temperature Range | |
@ 20ºC | -38C to 160 C |
@ 0ºC | -50 to 160°C |
Stability | |
As Liquid Bath | ± 0.01 °C |
As Dry block | ± 0.01 °C |
Uniformity as Liquid Bath | |
Axial | ± 0.02 °C @ -10 °C ± 0.02 °C@ 0°C ± 0.01 °C @ 120 °C |
Radial | ± 0.01 °C @ -10 °C ± 0.01 °C@ 0°C ± 0.02 °C @ 120 °C |
Accuracy | |
with Internal Sensor | ± 0.15 °C |
with External Sensor SRS (*) | ± 0.08 °C  ( TC version only) |
Display | |
Set Point | -50 to 160°C |
Resolution | user-selectable 0.1 , 0.01 C |
Units | user-selectable C, F,K |
Calibration Volume | |
As Liquid Bath | 44.5 x 185 mm |
As dry block | 45 x 170 mm ( dry inserts optional) |
Heating time | ambient to 120°C: 25 min ( include time to stability) |
Cooling time | ambient to -10°C : 25 min (include time to stability) |
Thermocouple Input (TC version) | |
Type | K,J,N,R,S,E,T,B |
Range | 0 to 1300 °C |
Accuracy | ± 1°C Include CJC error |
Resolution | user-selectable 0.1 , 0.01 °C |
RTD’S input ( TC version only) | |
Type | Pt 100 (3/4 wires) |
Range | -100 C to 700 °C |
Accuracy | ± 0.3 °C |
Resolution | user-selectable 0.1 , 0.01 °C |
General | |
Computer interface | RS 232 as standard ( UBS adapter supplied as option) and EIUK software included |
Switch Test | Yes |
Ramp | Yes fully programmable |
Weight | 10.2 Kg |
Dimensions | 160x3350x365 mm |
Power Requirement | 115 o 230 V – 50/60 Hz |
Power | 300 VA |
 Notes | * Require SRS sensor and special UKAS matching calibration. Long term drift not included |
Specifications
Eurotron PLB 157 Specifications
Temperature Range | |
@ 20ºC | -38C to 160 C |
@ 0ºC | -50 to 160°C |
Stability | |
As Liquid Bath | ± 0.01 °C |
As Dry block | ± 0.01 °C |
Uniformity as Liquid Bath | |
Axial | ± 0.02 °C @ -10 °C ± 0.02 °C@ 0°C ± 0.01 °C @ 120 °C |
Radial | ± 0.01 °C @ -10 °C ± 0.01 °C@ 0°C ± 0.02 °C @ 120 °C |
Accuracy | |
with Internal Sensor | ± 0.15 °C |
with External Sensor SRS (*) | ± 0.08 °C  ( TC version only) |
Display | |
Set Point | -50 to 160°C |
Resolution | user-selectable 0.1 , 0.01 C |
Units | user-selectable C, F,K |
Calibration Volume | |
As Liquid Bath | 44.5 x 185 mm |
As dry block | 45 x 170 mm ( dry inserts optional) |
Heating time | ambient to 120°C: 25 min ( include time to stability) |
Cooling time | ambient to -10°C : 25 min (include time to stability) |
Thermocouple Input (TC version) | |
Type | K,J,N,R,S,E,T,B |
Range | 0 to 1300 °C |
Accuracy | ± 1°C Include CJC error |
Resolution | user-selectable 0.1 , 0.01 °C |
RTD’S input ( TC version only) | |
Type | Pt 100 (3/4 wires) |
Range | -100 C to 700 °C |
Accuracy | ± 0.3 °C |
Resolution | user-selectable 0.1 , 0.01 °C |
General | |
Computer interface | RS 232 as standard ( UBS adapter supplied as option) and EIUK software included |
Switch Test | Yes |
Ramp | Yes fully programmable |
Weight | 10.2 Kg |
Dimensions | 160x3350x365 mm |
Power Requirement | 115 o 230 V – 50/60 Hz |
Power | 300 VA |
 Notes | * Require SRS sensor and special UKAS matching calibration. Long term drift not included |