GW Instek GPM 8213 Single Phase Digital Power Meter
The GW Instek GPM 8213 is a important apparatus for engineers, professionals, and analysts working with single-phase AC control supplies. It gives them with the vital estimations and investigation capabilities to assess the execution, productivity, and quality of their control supply frameworks. Whether it’s for vitality reviews, control utilization examination, or control quality checking, the GPM-8213 offers a dependable and precise arrangement for single-phase control estimations.
- In-stock items are delivered within 2 business days.
- We are not Supporting COD Payments Terms Outside UAE
- For out-of-stock items, delivery may take 2 to 4 business weeks.
Order Via phone number
GW Instek GPM 8213 Single-Phase Digital Power Meter
The GW Instek GPM 8213 is a single-phase computerized control meter outlined particularly for AC control estimations in single-phase (1P/2W) AC control supply applications. This control meter offers exact and solid estimations of power-related parameters for single-phase frameworks. With its specialized plan, it gives clients the capability to degree and analyzes different angles of control, empowering them to screen and assess the execution of their AC control supply.
The GW Instek GPM 8213 highlights a user-friendly interface and a five-digit estimation shown on a 5-inch TFT LCD screen, permitting clients to peruse and decipher the estimation effectively that comes about. It offers a wide run of control estimation parameters, giving comprehensive data about the control characteristics of the single-phase AC supply. These parameters incorporate voltage, current, control calculation, dynamic control, receptive control, and clear control.
With a high-precision estimation capability and an inspecting rate of 300 kHz, the GW Instek GPM 8213 guarantees exact and solid estimations indeed in energetic control conditions. It offers a test transmission capacity of 0.1 Hz to 100 kHz, permitting clients to analyze control changes and waveform characteristics in a wide recurrence extent. This empowers clients to pick up experience in the quality and proficiency of their single-phase AC control supply.
The GW Instek GPM-8213 is an important apparatus for engineers, professionals, and analysts working with single-phase AC control supplies. It provides them with vital estimations and investigation capabilities to assess the execution, productivity, and quality of their controlled supply frameworks. Whether it’s for vitality reviews, control utilization examination, or control quality checking, the GPM-8213 offers a dependable and precise arrangement for single-phase control estimations.
GW Instek GPM 8213 Single Phase Digital Power Meter specifications:
- 4” TFT LCD
- Two data display modes
- Standard: two major measurement items + six secondary measurement items
- Simple: displaying test data of four different measurement items
- Met the requirements for IEC 62301 power measurement
- Voltage/ current test frequency bandwidth: DC~6kHz
- Watt Resolution: 1 mW
- Lowest current range: 5 mA, resolution: 0.1 uA
- Current/voltage measurements reach CF=3 for distorted wave and CF=6 for half range
- W-h (power vs time)/A-h (current vs time) integration function
- Total harmonic distortion measurement
- Front panel test terminal
- Standard interface: RS-232C, USB device, LAN
- Optional interface: GPIB
- Optional test fixture: GPM-001
specifications
| Item Spec. | |
|---|---|
| Input voltage | 600 Vrms |
| Input current | 20 Arms |
| Input impedance (50/60 Hz) — Voltage | 2.4 MΩ |
| Input impedance (50/60 Hz) — Current | 5 mA – 200 mA: 500 mΩ 0.5 A – 20 A: 5 mΩ |
| Maximum display voltage | 700 Vrms |
| Maximum display current | 25 Arms |
| Maximum allowable isolation voltage | 300 V |
| Low frequency filter | Cut-off frequency 500 Hz |
| Display / Processing | |
| Synchronization frequency | 45 Hz ∼ 6 kHz |
| Refresh rate | 10 times/sec |
| Average | 1, 2, 4, 8, 16, 32, 64 |
| Displayed mode | Standard: 8 items simultaneously. Simple: 4 items simultaneously. |
| Displayed digits | 5 |
| Voltage converter (PT ratio) | 1 to 9,999.999 |
| Current converter (CT ratio) | 1 to 9,999.999 |
| Measurement items | |
| Voltage | Vdc, Vrms, V+pk, V-pk |
| Current | Idc, Irms, I+pk, I-pk |
| Power | P, P+pk, P-pk, VA, VAR |
| Factors | Crest Factor: CFV, CFI | Power Factor: PF |
| Frequency | VHz, IHz |
| Angle | DEG |
| Total Harmonic Distortion | THDV, THDI |
| Integration | Time, WP, WP+, WP-, q, q+ |
| Voltage Measurement | |
| Measurement range | CF = 3: 15 V, 30 V, 60 V, 150 V, 300 V, 600 V CF = 6: 7.5 V, 15 V, 30 V, 75 V, 150 V, 300 V |
| Crest factor | 3, 6 |
| Accuracy | Effective range: 1% to 105% of range DC: ±(0.2% reading + 0.2% range) 45 Hz ≤ f ≤ 66 Hz: ±(0.1% reading + 0.1% range) 66 Hz < f ≤ 1 kHz: ±(0.1% reading + 0.2% range) 1 kHz < f ≤ 6 kHz: ±3% of range Filter ON: Increase 0.3% reading @ 45–66 Hz |
| Temperature effect | 5–18 °C / 28–40 °C: Increase ±0.03% reading / °C |
| Residual noise | 0.5% of range |
| Current Measurement | |
| Measurement range | CF = 3: 5 mA, 10 mA, 20 mA, 50 mA, 100 mA, 200 mA, 500 mA, 1 A, 2 A, 5 A, 10 A, 20 A CF = 6: 2.5 mA, 5 mA, 10 mA, 25 mA, 50 mA, 100 mA, 250 mA, 0.5 A, 1 A, 2.5 A, 5 A, 10 A |
| Crest factor | 3, 6 |
| Accuracy | Effective range: 1% to 105% of range DC: ±(0.2% reading + 0.2% range) 45 Hz ≤ f ≤ 66 Hz: ±(0.1% reading + 0.1% range) 66 Hz < f ≤ 1 kHz: ±(0.1% reading + 0.2% range) 1 kHz < f ≤ 6 kHz: ±3% of range Filter ON: Increase 0.3% reading @ 45–66 Hz |
| Temperature effect | 5–18 °C / 28–40 °C: Increase ±0.03% reading / °C |
| Residual noise | 0.5% of range |
| Power Measurement | |
| Accuracy | Effective range: 1% to 110% of range DC: ±(0.2% reading + 0.2% range) 45 Hz ≤ f ≤ 66 Hz: ±(0.1% reading + 0.1% range) 66 Hz < f ≤ 1 kHz: ±(0.1% reading + 0.3% range) 1 kHz < f ≤ 6 kHz: ±3% of range Filter ON: Increase 0.3% reading @ 45–66 Hz Temperature effect (5–18 °C / 28–40 °C): Increase ±0.03% reading / °C |
| Frequency Measurement | |
| Measurement range | Filter ON: 30.000 Hz to 499.99 Hz Filter OFF: 30.000 Hz to 9.9999 kHz |
| Measurement items | Voltage, Current |
| Effective input range | 10% to 105% of voltage input range |
| Accuracy | ±0.06% reading |
| Integrator Measurement | |
| Integrator accuracy | ±(Accuracy of voltage or current + 0.1% reading) |
| Time range | 0 h 0 min to 9999 h 59 min |
| Time accuracy | ±0.01% ±1 s |
| General | |
| Display | 4″ TFT LCD |
| Interfaces | Standard: RS-232C, USB device, LAN | Optional: GPIB |
| Power source | AC 100–240 V, 50–60 Hz |
| Power consumption | 25 VA max. |
| Dimensions & weight | 270 (W) × 110 (H) × 350 (D) mm, Approx. 2.9 kg |